388 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
error message Initialization Error (Tcfs temperature not settled, error value)
description Temperature has not settled at the tuning control feedback system
error value 0 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (TmpCalDataInit, error value)
description Error accessing caldata
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (UpPiezoEnable, error value)
description Over current
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error - module temperature out of range
(please wait a little bit and try again)!
description Module temperature is too high
error value (no error value)
action Doublecheck whether the fan is working and not blocked. Wait some time and reboot the
instrument. If the failure persists, please return the unit to your local service center for repair.