392 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
error value 23071 test laser current not in limits
23072 current laser safety did not trigger at overcurrent
action Please return the unit to your local service center for repair
error message Main Piezo Amp. Selftest Failed (error value)
description Verifies if the piezo amplifier reaches the upper and lower limit. An error may indicate a defect amplifier
transistor
error value 23091 piezo amplifier could not set to maximum voltage
23092 piezo amplifier could not set to minimum voltage
action Please return the unit to your local service center for repair
error message Main Cavity Motor Selftest Failed (error value)
description Verifies the functionality of the cavity motor
error value 23081 indicates an open motor connector
23082 indicates that a motor coil is defect
23083 indicates a short in a motor coil
action Please return the unit to your local service center for repair
error message TCFS DC Selftest Failed (error value)
description Tests all fixed DC voltages available at the TCFS board mux
error value 25011 +10V measurement voltage on TCFS board not in limits
25012 -10V measurement voltage on TCFS board not in limits
25013 +5V reference voltage on TCFS board not in limits
25014 -5V reference voltage on TCFS board not in limits
25015 +2.5V reference voltage on TCFS board not in limits

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 8163A and is the answer not in the manual?

Keysight Technologies 8163A Specifications

General IconGeneral
BrandKeysight Technologies
Model8163A
CategoryMultimeter
LanguageEnglish

Related product manuals