404 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
error message E1x060: 0.25V offset failed (a, b)
description Verifies that the offset voltage for the 16-bit ADC (250 mv) is correctly added to the
input signal
test value a shows the ADC reading
b shows the maximum difference to the exact ADC reading of 3277
x refers to the module channel tested
action Please return the unit to your local service center for repair
error message E1x061: PGA DAC adj failed (a, b, c)
description Tries to adjust the offset voltage of the PGA
test value a shows the ADC offset reading
b shows the limit for an upper offset
c shows the limit for a lower offset
x refers to the module channel tested
action Please return the unit to your local service center for repair