EasyManua.ls Logo

Keysight Technologies 8163A - Page 405

Keysight Technologies 8163A
466 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
404 8163A/B, 8164A/B, 8166A/B User’s Guide
14 Error Messages
error message E1x060: 0.25V offset failed (a, b)
description Verifies that the offset voltage for the 16-bit ADC (250 mv) is correctly added to the
input signal
test value a shows the ADC reading
b shows the maximum difference to the exact ADC reading of 3277
x refers to the module channel tested
action Please return the unit to your local service center for repair
error message E1x061: PGA DAC adj failed (a, b, c)
description Tries to adjust the offset voltage of the PGA
test value a shows the ADC offset reading
b shows the limit for an upper offset
c shows the limit for a lower offset
x refers to the module channel tested
action Please return the unit to your local service center for repair

Table of Contents