104 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimizing Performance for All Models
Using Flatness Correction
c. Open the User Flatness table editor and preset the cal array:
Press Return > Configure Cal Array > More > Preset List > Confirm Preset with Defaults.
d. In the Step Array menu, enter the desired flatness–corrected start and stop frequencies,
and the number of points:
e. Populate the user flatness correction array with the step array configured in the previous
step:
Press Return > Load Cal Array From Step Array > Confirm Load From Step Data.
f. Set the output amplitude to 0 dBm.
g. Turn on the RF output.
Press
More > Configure Step
Array >
Freq Start > 500 > MHz >
Freq Stop > 1 > GHz >
# of Points > 10 > Enter