Keysight EXG and MXG X-Series Signal Generators User’s Guide 71
Front Panel Operation for All Models
Using Analog Modulation on the RF Carrier Signal (Option UNT)
Simultaneous Modulation
The Keysight X-Series signal generators are capable of simultaneous
modulation. All modulation types (AM, FM, φM, Pulse, and I/Q) may be
simultaneously enabled, but there are some exceptions. Refer to Table 4-1.
Table 4-1 Simultaneous Modulation Type Combinations
AM
a
a. Linear AM and Exponential AM cannot be enabled simultaneously. Refer to Chapter 5.
FM φM
Pulse
b
b. Pulse modulation requires Option UNW. Refer to Chapter 6.
I/Q
AM--xxxx
FM
x
c
c. FM and φM cannot be enabled simultaneously.
-- not
applicable
xx
φM
x
c
not
applicable
-- x x
Pulsexxx--x
I/Qxxxx--