Keysight EXG and MXG X-Series Signal Generators User’s Guide 127
Optimizing Performance for All Models
Using an Output Offset, Reference, or Multiplier
To set a new frequency or amplitude reference, turn the frequency reference off, and then follow the
steps above.
Setting a Frequency Multiplier
Using a frequency multiplier, the signal generator can display a frequency that is the multiple
(positive or negative) of the output value.
Displayed Value = multiplier value ´ output frequency
Output Frequency = displayed value ¸ multiplier value
To set a frequency multiplier:
1. Press Frequency > Freq Multiplier > multiplier value > x.
2. Set the desired frequency.
The display equals the output frequency times the multiplier value.
Examples
Parameter
Example
#1
Example
#2
Example
#3
Comments
Frequency Multiplier: 3 3 4 The multiplier range can be set from:
+0.001 to +1000
–1000 to –0.001
Entered (and displayed)
Value:
600 MHz 600
MHz
8 GHz
Output Frequency: 200 MHz 200 MHz 2 GHz
The signal generator alerts you if the output frequency is
out of range.
Indicates that a frequency multiplier is on

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals