Keysight EXG and MXG X-Series Signal Generators User’s Guide 393
Using Avionics VOR/ILS Softkeys for N5171B/72B and N5181B/82B with Option 302
Using ILS Marker Beacon Softkeys
To set the ILS Marker Beacon AM depth
1. Press Aux Fctn > Avionics > Marker Beacon
2. Press Marker Depth
(Default <value> is 95%.)
3. Select a <value> from 0 to 99.9% and press Enter.
These softkeys set the AM depth on the Marker Beacon carrier.
To set the ILS Marker Beacon inner marker frequency
1. Press Aux Fctn > Avionics > Marker Beacon
2. Press Marker Beacon
3. Press Inner
4. Press Marker Freq
(Default <value> is 3000 Hz.)
5. Select a <value> in Hz | kHz | MHz | GHz from 0 Hz to 10 MHz.
These softkeys set the frequency for the Inner Marker Beacon.
To set the ILS Marker Beacon middle marker frequency
1. Press Aux Fctn > Avionics > Marker Beacon
2. Press Marker Beacon
3. Press Middle
4. Press Marker Freq
(Default <value> is 1300 Hz.)
5. Select a <value> in Hz | kHz | MHz | GHz from 0 Hz to 10 MHz.
These softkeys set the frequency for the Middle Marker Beacon.
5=74.700 16=74.975 27=75.250
6=74.725 17=75.000 28=75.275
7=74.750 18=75.025 29=75.300
8=74.775 19=75.050 30=75.325
9=74.800 20=75.075 31=75.350
10=74.825 21=75.100 32=75.375
11=74.850 22=75.125 33=75.400

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals