Keysight EXG and MXG X-Series Signal Generators User’s Guide 69
Front Panel Operation for All Models
Using Swept Signal Mode (Configuring a Swept Output)
Example: Manual Control of Sweep
1. Set up either a step sweep (page 64) or a list sweep (page 66).
2. In the Sweep/List menu, select a parameter to sweep: Press Sweep > parameter > Return.
3. Select manual mode: Press More > Manual Mode Off On.
When you select manual mode, the current sweep point becomes the selected manual point.
4. If it is not already on, turn the RF output on: Press RF On/Off.
5. Select the desired point to output: Press Manual Point > number > Enter.
The progress bar changes to indicate the selected point.
6. Use the knob or arrow keys to move from point to point. As you select each point, the RF
output changes to the settings in that selection.
The parameters of the selected sweep point (point 3
of 6 in this example) define the signal available at the
RF Output connector.
When you enter a manual point,
the progress bar moves to and
stops at the selected point.
The SWMAN annunciator
indicates that the sweep
is in manual mode.
When you turn manual mode on, the
current sweep point becomes the
manual point.
The currently sweeping
parameter stops and
displays the selected point.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals