188 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Digital Operation for N5172B/82B with Options 653/655/656/657
Triggering a Waveform
Example: Segment Advance Triggering
Segment advance triggering enables you to control the segment playback within a waveform
sequence. This type of triggering ignores the repetition value (page 160). For example if a segment
has repetition value of 50 and you select Single as the segment advance triggering mode, the
segment still plays only once. The following example uses a waveform sequence that has two
segments.
If you have not created and stored a waveform sequence, refer to “Creating a Waveform Sequence”
on page 159.
1. Preset the signal generator.
2. Configure the RF output:
— Set the desired frequency.
— Set the desired amplitude.
— Turn on the RF output.
3. Select a waveform sequence for playback:
a. Press Mode > Dual ARB > Select Waveform.
b. In the Sequence On column, highlight a waveform sequence file.
c. Press Select Waveform.
4. Set the triggering as follows:
— Trigger Type: continuous Segment Advance
Press Trigger Type > Segment Advance > Continuous.
— Trigger source: Trigger key
Press Trigger Source > Trigger Key.
5. Generate the waveform sequence:
Press ARB Off On until On highlights.
6. (Optional) Monitor the waveform:
Connect the RF OUTPUT of the signal generator to the input of an oscilloscope, and configure
the oscilloscope so that you can see the signal.
7. Trigger the first waveform segment to begin playing continuously:
Press the Trigger key.
8. Trigger the second segment:
Press the Trigger key.
Pressing the Trigger key causes the currently playing segment to finish and the next segment
to start.
If the last segment in the sequence is playing, pressing the Trigger key causes the first
segment in the waveform sequence to start when the last segment finishes.