Keysight EXG and MXG X-Series Signal Generators User’s Guide 113
Optimizing Performance for All Models
Using I/Q Mod Skew Cal
SA Config Softkeys
For details on each key, use key help as
described on page 56.
SCPI Commands:
:SYSTem:SANalyzer:COMMunicate:TYPE SOCKets|VXI11
:SYSTem:SANalyzer:COMMunicate:LAN:DEVice
<"deviceName">
:SYSTem:SANalyzer:COMMunicate:LAN:DEVice?
:SYSTem:SANalyzer:COMMunicate:LAN:IP <"ipAddr">
The status window
Attempts to connect to the
specified external spectrum
analyzer and execute a "*IDN?"
SCPI command. If the result is
"Connected, but no '*IDN?'
response", then the IP address
connected to something, but the
socket port or VXI-11 device
name was not correct.
Recognized Spectrum Analyzer
models are currently: E4440A,
E4443A, E4445A, E4446A,
E4448A, and N9020A.
I/Q > More > Int Channel Correction > Configure
Sets the spectrum analyzer
connection type to Sockets (LAN).
Sets the spectrum analyzer
connection type to VXI-11 (LAN).
This connection type can also be
used to connect to a GPIB
spectrum analyzer via a
LAN-to-GPIB gateway.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals