396 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Using Avionics VOR/ILS Softkeys for N5171B/72B and N5181B/82B with Option 302
Using ILS Marker Beacon Softkeys
c. Press Marker Freq
d. (Optional) Select a <value> from 0 Hz to 10 MHz.
8. Set the modulation to on.
a. Press MOD On and verify that the front panel LED is illuminated,
indicating that it is on.
(Default is on).
9. Set a desired RF power level.
a. Press Amptd
(Default power level is around -144 dBm, and depends on the signal
generator model being used, so the signal may be very low and hard
to find if you do not specify a high enough power level.)
b. Select a <value> from -20 dBm to -40 dBm.
10.Set the RF output to on.
a. Press RF On and verify that the front panel LED is illuminated,
indicating that it is on.
(Default is off).
11.Verify that all settings are as expected on the signal generator display.
a. Press Aux Fctn > Avionics > Marker Beacon
b. Review the settings under
Marker Beacon Status Information