176 Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
Config Key Group
Gain and Offset Parameters for calculating the wait time. See the following
formula.
If State = ON and Automatic = ON:
wait time = Gain initial wait time Offset
If State = ON and Automatic = OFF:
wait time = Offset
If State = OFF:
wait time = 0
The initial wait time is automatically set by the instrument, and cannot be
changed.