132 Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
Hard Keys and Rotary Knob
Hard Keys and Rotary Knob
Standby switch Turns the instrument on or off. During the power is on, the LED below the switch
turns green. For the B2983B/B2987B, the LED turns orange during the switch is
off and the battery charging is in progress.
Single Starts a single (one shot) measurement or initiates trigger system. Also triggers
the staircase sweep output or square waveform output of B2985B/B2987B. If a
repeat measurement is in progress, it stops it and starts a single measurement.
The measurement result is displayed on the Meter view, Graph view, Histogram
view, or Roll view.
Once a single measurement starts, the data buffer is cleared and stores the last
single measurement result (maximum 100000 data). It can be listed and displayed
on the Measure Result dialog box.
If measurement is not performed properly, check the trigger setting. The trigger
type must be set to AUTO, or the trigger count (Count) must be set properly. See
“Trigger Setup Sub-panel” on page 157.
Run/Stop Starts a repeat (continuous) measurement. If a repeat measurement is in progress,
this stops it. The measurement result is displayed on the Meter view, Histogram
view, or Roll view.
Once a repeat measurement starts, the data buffer is cleared and stores the last
repeat measurement result (maximum 1000 data). It can be listed and displayed
on the Measure Result dialog box.
The repeat measurement is performed with the following trigger setup. Settings of
“Trigger Setup Sub-panel are ignored.
Acquire trigger (measurement trigger): Initiate
ARM acquire count: Infinite
ARM acquire source: AUTO (automatic internal)
TRIGger acquire count (Measure Count): 100
TRIGger acquire source (Measure Trigger): AUTO (automatic internal)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight B2980B Series and is the answer not in the manual?

Keysight B2980B Series Specifications

General IconGeneral
BrandKeysight
ModelB2980B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals