EasyManuals Logo

Keysight B2980B Series User Manual

Keysight B2980B Series
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #254 background imageLoading...
Page #254 background image
254 Keysight B2980B User’s Guide, Edition 1
Function Details
Initial Settings
Initial Settings
The B2980 is initialized by turning it on or resetting it (by the *RST command or
the device clear). Initial settings of the B2980 are shown in this section.
Table 5-8 System Initial Settings
Setup item Power on Reset
GPIO function Digital In
GPIO polarity Negative
GPIO output trigger type Edge
GPIO output trigger timing Both
GPIO output trigger pulse width 100 s
BNC trigger input polarity Negative
BNC trigger output polarity Negative
BNC trigger output type Edge
BNC trigger output timing Both
BNC trigger output pulse width 100 s
Byte order (measurement data) Normal (Data Swap: OFF)
Data format (measurement data) ASCII
Data format (GPIO data) ASCII
Data format (status register) ASCII
Data elements (measurement data) All (voltage, current, resistance, voltage source output, time,
temperature, humidity, status)
Data elements (calculate) CALC
Automatic timestamp reset ON
Mass storage directory (Root)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight B2980B Series and is the answer not in the manual?

Keysight B2980B Series Specifications

General IconGeneral
BrandKeysight
ModelB2980B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals