Keysight B2980B User’s Guide, Edition 1 171
Front Panel Reference
Config Key Group
Config Key Group
Press the Config key to display the following softkeys for setting several functions
of the meter and the voltage source.
Source (B2985B/B2987B) Displays the following softkeys for setting the voltage source
operation.
Connection Source operation and connection setup. See “Output
Connection Dialog Box” on page 173.
Sweep Sweep source detail setup. See “Sweep Dialog Box” on
page 174.
Measure Displays the following softkeys for setting the measurement operation.
Connection (B2985B/B2987B) Measurement operation and connection
setup. See “Input Connection Dialog Box” on page 174.
Filter Digital filter setup for measurement. See “Measure Filter Dialog
Box” on page 175.
Ranging Automatic ranging speed setup, Normal or Fast
Normal Performs the basic auto-ranging operation.
The channel automatically sets the range
which provides the best resolution in
performing the measurement.
Fast For performing high throughput
measurement, reduces observation time to
determine measurement range. This setup
can be used when the measurement signal
noise (superposed from AC power, etc.) is
low.
The softkey displays the present Normal/Fast status of this
function. Pressing this key switches the status.
Coulomb (B2985B/B2987B) Displays the following softkeys for setting
the automatic discharge function of the charge measurement.