208 Keysight B2980B User’s Guide, Edition 1
Function Details
Measurement Time
Overhead Time
Overhead time is the time required for range change, etc. This time depends on
the measurement condition, and cannot be specified. A major element of the
overhead time is range change time during measurement (when measurement
ranging mode is set to AUTO).
To Control Source/Measure Timing (B2985B/B2987B)
The voltage source output and measurement timing can be controlled by the
following parameters. See Figure 5-2. This figure shows an example of sweep
output. For a bias output, focus the shape of a sweep step only.
1. Source delay
The source delay time is defined as the time from trigger to start of a voltage
source output.
2. Measure delay
The measurement delay time is defined as the time from trigger to start of a
measurement.
3. Period
The period is the trigger interval. It can be defined for the TIMER or MANUAL
trigger type (trigger source) and for the voltage source output and
measurement actions individually.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight B2980B Series and is the answer not in the manual?

Keysight B2980B Series Specifications

General IconGeneral
BrandKeysight
ModelB2980B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals