Keysight B2980B User’s Guide, Edition 1 85
How to Perform Measurement
Current Measurement
Current Measurement
The B2981B/B2983B/B2985B/B2987B supports the current measurement
capability shown in Table 3-1.
Table 3-1 Current Measurement Range, Value, and Resolution
Requirements Before turning the instrument on, connect cable, test leads, test fixture, and so on,
used for the measurement. See Figures 3-2 and 3-3 for connection examples.
The following accessories can be used.
Triaxial cable, 200 V, 1.5 m
Triaxial bulkhead connector, if needed
Banana to lug cable, for connecting Common to chassis ground
Instead of the triaxial cable and the triaxial bulkhead connector, Keysight N1415A
Triaxial to alligator clip cable, 200 V, 1.5 m can be used.
Range value Measurement value Display resolution
2 pA 0 |I| 2.1 pA 1 aA
20 pA 0 |I| 21 pA 10 aA
200 pA 0 |I| 210 pA 100 aA
2 nA 0 |I| 2.1 nA 1 fA
20 nA 0 |I| 21 nA 10 fA
200 nA 0 |I| 210 nA 100 fA
2 A0 |I| 2.1 A1 pA
20 A0 |I| 21 A10 pA
200 A0 |I| 210 A100 pA
2 mA 0 |I| 2.1 mA 1 nA
20 mA 0 |I| 21 mA 10 nA

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight B2980B Series and is the answer not in the manual?

Keysight B2980B Series Specifications

General IconGeneral
BrandKeysight
ModelB2980B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals