Keysight B2980B User’s Guide, Edition 1 239
Function Details
Trigger System
Trigger Source
B2980 supports the following trigger sources. The trigger source must be set to
the ARM event and the TRIGGER event individually.
AUTO (automatic internal, AINT): Trigger source best suited for the present
operating mode is automatically selected by the internal algorithms.
BUS: Remote interface trigger command such as GET, TRIGger, and *TRG
TIMER (timer, TIMer): Signal internally generated every interval set by the
Period field
INT1 or INT2 (internal): Signal from the internal bus 1 or 2
EXT1, EXT2, EXT3, EXT4, EXT5, EXT6, or EXT7 (external): Signal from the DIO
pin n, which is an I/O port of the Digital I/O D-sub connector on the rear panel.
n = 1 to 7
LAN: LXI trigger
TIN: Signal from the Trigger Input connector
Device Actions
B2980 supports the following device actions. Also see Figure 5-11.
Transient device action (voltage source output for B2985B/B2987B)
Voltage source channel starts voltage output for applying one output value if
the TRIGGER event was detected and the Source delay time elapsed.
Acquire device action (measurement)
Measurement channel starts measurement for obtaining one measurement
data if the TRIGGER event was detected and the Measure delay time elapsed.