198 Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
Display Key Group
Restore Trigger Mode
on Local Stores the trigger mode setting on entering GP-IB remote
state. ON or OFF (default)
ON: Saves the trigger mode setting in the nonvolatile memory
on entering GPIB remote state and restores the saved trigger
mode setting when returning to GPIB local mode.
OFF: Disables this function.
Measured Value
Format
(Meter View) Selects the display format of measured value to be displayed
on Meter view. ENG (default) or EXP
ENG: Displays the measured value in engineering format.
EXP: Displays the measured value in exponential format.