234 Keysight B2980B User’s Guide, Edition 1
Function Details
Trace Buffer
• BIN: Bin number
The statistical data (except for PKPK and BIN) of CURR, CHAR, VOLT, or RES data
can be displayed on the Trace Statistical Result dialog box. See “Trace Statistical
Result Dialog Box” on page 186 for displaying the data.
Use an external computer to display the data which cannot be displayed on the
B2980 screen. The data can be saved in a USB flash drive connected to the front
panel USB-A connector or can be read by using SCPI commands.
Do not forget to save or read the data before turning the B2980 off. The trace
buffer is cleared by turning the instrument off.