Keysight B2980B User’s Guide, Edition 1 241
Function Details
Trigger System
Table 5-7 Trigger Output Timing and Setup Parameters
For using a digital I/O pin, it is necessary to set function of the pin, output trigger
polarity, type, pulse width, and the Output Trigger Timing parameter. See “DIO
Configuration Dialog Box” on page 194 and “Using Digital I/O and Trigger In/Out”
on page 78.
For using the Trigger Out connector, it is necessary to set output trigger polarity,
type, pulse width, and so on. See “BNC Configuration Dialog Box” on page 195.
Trigger output timing Layer Action
Trigger
Output
Output
Trigger
Timing
Starting ARM loop ARM TRANS. for
transient
action or ACQ.
for acquire
action
ON BEFORE
Ending ARM loop ARM ON AFTER
Starting TRIGGER loop TRIGGER ON BEFORE
Ending TRIGGER loop TRIGGER ON AFTER
Starting Transient device action ACTION TRANS. ON BEFORE
Ending Transient device action ACTION TRANS. ON AFTER
Starting Acquire device action ACTION ACQ. ON BEFORE
Ending Acquire device action ACTION ACQ. ON AFTER

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight B2980B Series and is the answer not in the manual?

Keysight B2980B Series Specifications

General IconGeneral
BrandKeysight
ModelB2980B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals