116
TOS3200
IEEE488.2 Common Commands
*OPC
Section 12.5.3 in
IEEE 488.2-1992
Sets the OPC bit (bit 0) of the event status register when the processing of all commands
standing by is complete.
Command *OPC
*OPC?
Response
Returns 1 when the processing of all commands standing by is complete.
*OPT
Queries the option (HP21-TOS only) that is installed in the TOS3200.
Command *OPT?
Response
If the HP21-TOS is installed, “HP21-TOS” is returned. Returns 0 if the option is not installed.
*RCL
Aborts the measurement operation and reads the contents stored in memory. The operation
is the same as MEM:RCL.
Command *RCL <NR1>
Parameter Value: 0 to 99 Memory number
An SCPI error (-222, “Data out of range”) occurs if outside the range.
*RST
p. 196
Aborts the measurement operation and initializes the TOS3200 to factory default condition.
For the commands that are affected by *RST, see “Default values”.
Command *RST
*SAV
p. 196
Stores the present TOS3200 settings to memory. The operation is the same as MEM:SAV.
The settings that are stored are the same as those recalled using the *RCL command. See
“Default values”.
Command *SAV <NR1>
Parameter Value: 0 to 99 Memory number
An SCPI error (-222, “Data out of range”) occurs if outside the range.
*SRE
Sets the service request enable register.
The service request enable register is used to select the summary messages in the status
byte register that will be able to perform service requests.
To clear the service request enable register, send *SRE 0. If the register is cleared, service
requests cannot be generated by status information.