TOS3200
115
8
Remote Control
IEEE488.2 Common Commands
*CLS
Clears all event registers including the status byte, event status, and error queue.
p. 150
Command *CLS
*ESE
p. 153
Sets the event status register that is counted by the event summary bit (ESB) of the status
byte.
Command *ESE <NR1>
*ESE?
Parameter
Parameter Value: 0 to 255
An SCPI error (-222, “Data out of range”) occurs if outside the range.
(Example) When *ESE 16 is transmitted, bit 4 of the event status enable register is set. Each
time the execution error bit (bit 4) of the event status register is set, the summary
bit (ESB) of the status byte is set.
Response Returns the value of the event status enable register in the <NR1> form.
*ESR
p. 153
Queries the event status register. Registers that are read are cleared.
Command *ESR?
Response
Returns the value of the event status register in the <NR1> form and clears the register.
*IDN
Queries the model name, serial number, and firmware version of the TOS3200.
Command *IDN?
Response
The response to *IDN? is indicated below.
(Example) For TOS3200 with a serial number AB123456 and firmware version 1.00
Returns KIKUSUI,TOS3200,AB123456,1.00.
See