TOS3200
65
Starting the Test
4
TC and PCC Tests
Unable to Start the Test
READY is not displayed
You will not be able to start the test in the following conditions.
• When TC2/2 or PCC2/2 is displayed.
p. 68
• While recalling the panel memory or when the memory number has not been confirmed
while storing to the memory.
p. 96
• When the STOP switch is pressed (includes the condition in which the STOP signal is
being applied to the SIGNAL I/O connector).
p. 44
• When the power line of the EUT is turned on.
PASS or FAIL is displayed
You cannot start test while the judgment result is displayed. Press the STOP switch to set the
TOS3200 to READY status.
PROTECTION is blinking
p. 204
If “PROTECTION” is blinking on the screen, the TOS3200 is in the protection status. You
cannot start the test in this status. Eliminate the cause of the protection status, and then start
the test. For details, see “Protection Function”.
See
See
See
See
PCC 1 / 2
REMOTE PROTECTION
OFF ON
LOWER
UPPER TIMER
NORM
POL
NORM
COND
UPPER
ON
TIMER
30.0 mA
10
s
LOWER UPPER TIME/WAIT
Blinking