TOS3200
29
2
Installation and Preparation
Turning the Power On
Turning the POWER Switch On
1
Press the POWER switch to turn the power on ( ).
2
Check the firmware version (Ver x.xx) that appears on the screen.
p. 196
When you turn the POWER switch on for the first time, the TC measurement screen (TC1/2)
will appear after the firmware version (factory default setting).
Because the TOS3200 stores the conditions immediately before the POWER switch is turned
off, the TOS3200 starts up using the conditions that existed when the POWER switch was
turned off the last time.
System clock
The TOS3200 manages the calibration date using the internal system clock. If the preset cal-
ibration period is due when the power is turned on, the message “CAL DATE EXPIRED” will
appear on the screen.
p. 164
For the procedure to set the system clock and the remedy when the calibration period
expires, see “Time Settings and Calibration Management”.
TOS3200
LEAKAGE CURRENT TESTER
Ver 1.00
KIKUSUI ELECTRONICS CORP.
See
TC 1/2
READY
ENCPE OFF ON
PROBE LOWER
UPPER TIMER
POL
NORM
NORM
COND
UPPER
OFF
TIMER
30.0 mA
10
s
PROBE LOWER UPPER TIME/WAIT
See