TOS3200
141
Querying the Measured Value
8
Remote Control
READ:VOLT:UX / MEAS:VOLT:UX
p. 176
Queries the measured voltage between the measurement network reference points.
Command READ[:SCALar]:VOLTage:UX?
MEASure[:SCALar]:VOLTage:UX?
Response
Returns the measured voltage between reference points. The value is returned in the <NR3>
form.
Unit: V
READ:VOLT:UX :MAX / MEAS:VOLT:UX :MAX
Queries the maximum measured voltage between reference points that is retained on the
TOS3200. To clear the maximum measured voltage between reference points, send the
CURR:MAX:CLE command.
Command READ[:SCALar]:VOLTage:UX:MAXimum?
MEASure[:SCALar]:VOLTage:UX:MAXimum?
Response
Returns the maximum measured voltage between reference points that is retained on the
TOS3200. The value is returned in the <NR3> form. This query returns the same value as
READ:VOLT:UX/MEAS:VOLT:UX during the test wait time.
Unit: V
Querying the Measured Voltage between Measurement Ter-
minals A and B (Voltage Measurement Only)
READ:VOLT:AB / MEAS:VOLT:AB
Queries the measured voltage between measurement terminals A and B.
Command READ[:SCALar]:VOLTage:AB?
MEASure[:SCALar]:VOLTage:AB?
Response
Returns the measured voltage in the <NR3> form.
Unit: V
READ:VOLT:AB:MAX / MEAS:VOLT:AB:MAX
Queries the maximum measured voltage that is being held.
Command READ[:SCALar]:VOLTage:AB:MAXimum?
MEASure[:SCALar]:VOLTage:AB:MAXimum?
Response
Returns the maximum measured voltage that is held in the <NR3> form.
Unit: V
See