TOS3200
D
A List of Errors
Command errors
An error in the range [ -199, -100 ] indicates that an IEEE 488.2 syntax error has been
detected by the instrument’s parser. The occurrence of any error in this class shall cause the
command error bit (bit 5) in the event status register to be set.
Error code Error message description
-100 Command error Command error. A generic syntax error.
-101 Invalid character
Invalid character is present. A data element different from those that
are allowed was encountered.
-102 Syntax error
Syntax error. An unrecognized syntax was encountered in the
command string.
-103 Invalid separator An invalid separator exists in the command string.
-104 Data type error The parser recognized a data element different than one allowed.
-105 GET not allowed A Group Execute Trigger was received within a program message.
-108 Parameter not allowed More parameters were received than expected for the header.
-109 Missing parameter Fewer parameters were received than required for the header.
-110 Command header error An error was detected in the header.
-120 Numeric data error
Generated when parsing a data element which appears to be
numeric, including the nondecimal numeric types.
-130 Suffix error Generated when parsing a suffix.
-131 Invalid suffix
The suffix does not follow the syntax, or the suffix is inappropriate for
the TOS3200.
-134 Suffix too long The suffix of the numeric element is too long.
-138 Suffix not allowed
A suffix was encountered after a numeric element which does not
allow suffixes.
-140 Character data error Generated when parsing a character data element.
-141 Invalid character data
-144 Character data too Long Characters in the character data element are too long.
-148 Character data not allowed
-150 String data error Generated when parsing a string data element.
-160 Block data error Generated when parsing a block data element.
-170 Expression error Generated when parsing an expression data element.
-180 Macro error