TOS3200
Starting the Test
When a test starts (Cont’d)
The measured value is blinking
p. 82
If the measured value exceeds the measurement range specified by AUTO or FIX, the mea-
sured value will blink. See Table 6-2.
CONV is displayed
If the current conversion display function is enabled and the power is not applied to the AC
inlet for the EUT, “CONV
” will be displayed. The conversion display function is invalid in this
condition.
p. 200
If you are supplying the EUT power from an external source (not from the TOS3200) in the
TC measurement between two enclosures, turn the current conversion display function off.
SYSTEM1/5>CONV
Unable to Start the Test
READY is not displayed
You will not be able to start test in the following conditions.
When AUTO2/2 is displayed.
p. 96
When the STOP switch is pressed (includes the condition in which the STOP signal is
being applied to the SIGNAL I/O connector).
p. 44
When the power line of the EUT is turned on.
PROTECTION is blinking
p. 204
If “PROTECTION” is blinking on the screen, the TOS3200 is in the protection status. You will
not be able to start the test in this status. Eliminate the cause of the protection status, and
then start the test. For details, see “Protection Function”.
PASS or FAIL is displayed
You cannot start test while the judgment result is displayed. Press the STOP switch to set the
TOS3200 to READY status.
See
See
See
See
See
AUTO 1 / 2
00 TC•ENCPE––PNRM•NORM––– HOL
01 PCC––––––P
NRM•NORM––– 1s
–––END
PRG
01:TEST-1
TOTAL: 2
REMOTE PROTECTION
Blinking