EasyManua.ls Logo

Kikusui TOS3200 - Checking the EUT Operation

Kikusui TOS3200
208 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
TOS3200
Connecting the EUT
Checking the EUT Operation
The power is supplied through the power line for the EUT only during the test. To check the
EUT operation before the test, use the LINE ON key.
Fig. 4-4 LINE ON key
OVER LOAD PROTECTION indication
p. 204
If a current greater than or equal to 15.75 A flows through the power line of the EUT or the
power consumption of the EUT exceeds 1500 VA, the overload protection will be activated,
and the power line will be shut off. If this happens, the screen will display “OVER LOAD PRO-
TECTION,” and the TOS3200 will enter the protection status. To release the protection sta-
tus, disconnect the EUT, and press the STOP switch.
N
1500VA
MAX
L
G
DANGER
LEAKAGE CURRENT
AC LINE OUT
LINE
ON
EUT
A B
CAT II
REMOTE
This LED illuminates when the
power line to the EUT is on.
Press the LINE ON key to turn on the
power line to the EUT.
Press the key again to turn off the power
line.
Check the connection of the power line of the EUT before pressing the LINE ON key.
If the LINE ON LED is illuminated, the power for the EUT is supplied to the outlet on the front
panel and the AC LINE OUT terminal block on the rear panel.