196
TOS3200
F
A List of Default Settings
Initializing the TOS3200
While holding down the SHIFT key, turn
on the POWER switch. The TOS3200 dis-
plays the firmware version followed by
“INITIALIZING!” and starts to initialize.
If you initialize the TOS3200, various set-
tings such as the test conditions and the
saved data will be set to the default values
indicated below.
Default values
Item
Command
Setting after
initialization
Setting after sending the
command
*RST *RCL
Operation mode FUNC TC ←
Depends on the
memory contents.
TC
measurement
conditions
(TC)
PROBE TC:PROB ENCPE
←
Depends on the
memory contents.
POL TC:POL NORM
COND TC:COND NORM
LOWER TC:LIM:LOW 30 μA
LOWER ON/OFF TC:LIM:LOW:STAT OFF
UPPER TC:LIM:UPP 30 mA
UPPER ON/OFF TC:LIM:UPP:STAT ON
TIMER TC:TIM 10 s
TIMER ON/OFF TC:TIM:STAT OFF
WAIT TC:WAIT 1 s
WAIT ON/OFF TC:WAIT:STAT OFF
NTWK TC:NETW A
MODE TC:MODE RMS
RANGE TC:RANG:SEL AUTO
PCC
measurement
conditions
(PCC)
POL PCC:POL NORM
←
Depends on the
memory contents.
COND PCC:COND NORM
LOWER PCC:LIM:LOW 30 μA
LOWER ON/OFF PCC:LIM:LOW:STAT OFF
UPPER PCC:LIM:UPP 30 mA
UPPER ON/OFF PCC:LIM:UPP:STAT ON
TIMER PCC:TIM 10 s
TIMER ON/OFF PCC:TIM:STAT OFF
WAIT PCC:WAIT 1 s
WAIT ON/OFF PCC:WAIT:STAT OFF
MODE PCC:MODE RMS
RANGE PCC:RANGE:SEL AUTO