EasyManua.ls Logo

Kikusui TOS3200 - Other Settings

Kikusui TOS3200
208 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
148
TOS3200
Various Settings
Other Settings
SYST:ERR
Queries the oldest error information or event information from the error queue. The error
queue can store up to 255 errors.
The error queue is cleared completely using the *CLS command.
Command SYSTem:ERRor[:NEXT]?
Response
Returns the oldest error or event information in the error/event queue in response to
SYST:ERR? as follows:
(Example) If there is no error or event
Returns 0, "No error".
(Example) If a command that cannot be executed in the present operating condition is
received
Returns -221, "Settings conflict".
If multiple errors are present, errors are returned in order from the oldest error. The returned
errors are cleared.
SYST:KLOC
Sets or releases the panel operation lock.
Command SYSTem:KLOCk {ON|OFF|1|0}
SYSTem:KLOCk?
Parameter Value: ON (1) Lock the panel operation
OFF (0) Release the panel operation lock (default)
Response Returns the panel operation lock setting in the NR1 form.
SYST:LOC (RS232C and USB only)
Sets the TOS3200 operation to local mode (panel operation). This is a substitute command
for the IEEE488.1 REN message (Remote Disable).
SYST:REM or SYST:RWL is used to return to remote mode.
Command SYSTem:LOCal
SYST:OPT
Queries the option (HP21-TOS only) that is installed in the TOS3200. The operation is the
same as *OPT?.
Command SYSTem:OPTion?
Response
If the HP21-TOS is installed, “HP21-TOS” is returned. Returns 0 if the option is not installed.