Various Settings
8
Remote Control
SYST:REM (RS232C and USB only)
Sets the TOS3200 operation to remote mode. All panel keys except the LOCAL key and
STOP switch are locked. This is a substitute command for the IEEE488.1 REN message
(Remote Enable) and address designation.
SYST:LOC is used to return to local mode.
Command SYSTem:REMote
SYST:RWL (RS232C and USB only)
Sets the TOS3200 operation to remote mode. All panel keys except the STOP switch are
locked (LOCAL key is also locked). This is a substitute command for the IEEE488.1 llo mes-
sage (Local Lock Out).
SYST:LOC is used to return to local mode.
Command SYSTem:RWLock
SYST:VERS
Queries the version of the SCPI specifications to which the TOS3200 conforms.
Command SYSTem:VERSion?
Response
Always returns 1999.0 in response to SYST:VERS?.