GETTING STARTED MANUAL
WRXi-A-GSM-E Rev A 47
Starting TriggerScan
AfteryouhaveruntheTrainer,theTriggerListdisplaysalistofsmarttriggersetups.
Youcanaddorremovetriggersetups.Youcanalsoupdatetheselectedsmart
triggersetup.OnceyouhavemadeanychangestotheTriggerList,youarereadyto
startscanning.
1. TouchTriggerÆTriggerSet
upfromthemenubar.
2. OntheTriggerdialog,clicktheTriggerScantab.
3. Ifyouwanttoaddanewtriggersetup,touchtheTriggertabandsetthe
newtrigger.Then,touchtheAddNewbuttontoaddthenewtriggerto
theTriggerList.
4. Ifyo
uwanttodeleteatriggersetup,highlightthesetupintheTriggerList
andtouchtheDeleteSelectedbutton.
Note:IfyouwanttodeletealltriggersetupsintheTriggerList,touchtheDeleteAll
button.
5. Ifyouwanttoreplacetheselectedtriggersetupwiththecurrenttrigger
setup,highlightthesetupintheTriggerListandtouchtheUpdateSelected
button.
6. OnceyouhavemadeanychangestotheTriggerList,touchStartScan.The
oscilloscopeautomaticallysequencesthroughallthetriggersetups.
PLEASENOTETHEFOLLOWING:
Youcantunethedwelltimethatthescopewillwaitbeforeloadingthenext
triggersetupusingtheDwellTimedataentryfield.
IfyouhavePersistencedisplaymodeenabled,alltriggereventsarerecorded
onthedisplay.RefertothePersistenceSetuptopicforinstructionsonenabling
Persistencedisplaymode.
IfyouwantTriggerScantostopwhenthescopetriggersnext,checktheStop
OnTriggercheckbox.Youcanusethistoisolatespecifictriggersetups.
Saving TriggerScan Setups
YoushouldsaveTriggerScansetupsonceyouhavemadeanymodificationstothe
TriggerList.ThecurrentTriggerListwillnotbepreservedafterexitingthe
applicationunlessyousaveit.
OntheTriggerScandialog,touchinsidetheSetupFileNamedataentryfieldand
enterafilenameusingthepopupke
ypad.
OR

Table of Contents