WAVERUNNER Xi-A AND MXi-A SERIES
100 WRXi-A-GSM-E Rev A
Note:Therecoveryprocessproducesareplicaoftheoperatingsystemand
oscilloscopeapplicationsoftwareatthecurrentrevisionlevelswhenthe
oscilloscopewasmanufactured.Anyfurtherrevisionsoftheapplicationsoftware,
Windowsoperatingsystem,andvirusscandefinitionfilesarenotautomatically
upgraded.Therefore,aftercompletionofthediskimagerecovery,itishighly
recommendedtosearchvendorWebsitesandupgradetheindividualcomponents
tocurrentrevisionlevels.Thecurrentoscilloscopeapplicationsoftwarecanbe
downloadeddirectlyfromtheLeCroywebsiteatwww.lecroy.com.Sincethe
calibrationdatafortheoscilloscopeisstoredintheD:drive,currentcalibration
constantsarenotoverwrittenduringtherecoveryprocess.
RecoveryProcedure
1.
ConnectanetworkcabletotheLANportontherearofthe
oscilloscopeifyouintendtoactivatewindowsthroughtheInternet.
2.
Connectakeyboardandamousetotheoscilloscope.
3.
Applypowertotheoscilloscope.
4.
AssoonastheStartingAcronisLoader...PressF11forAcronis
St
artupRecoveryManagermessageappearsonthescreen,press
theF11keyuntiltherecoverysoftwarelogoappearsmomentarily.
TheAcronisTrueImageEchoWorkstationmainwindowis
displayed.
Note:DonotpressF11beforeyouseetheStartingAcronisLoader
messageoryouwillentertheMBbootsequenceselection.
5.
SelectRecoveryfromtheMainMenu.TheRestoreDataWizard
opens.
6.
ClickNextontheWelcomepage.
7.
OntheBackupArchiveSelectionpage,youchoosethezonewhere
therecoverypartitionislocated.ChooseAcronisSecureZone(this
iswheretherecoverydataislocatedonyouroscilloscope)an
dclick
Next.