WAVERUNNER Xi-A AND MXi-A SERIES
116 WRXi-A-GSM-E Rev A
Pass/Fail Testing
Testmultipleparametersagainstselectableparameterlimitsatthesametime.Pass
orfailconditionscaninitiateactionsincluding:documenttolocalornetworked
files,emailtheimageofthefailure,savewaveforms,sendapulseoutatthefront
panelauxiliaryBNCoutput,or(withGPIBoption)sendaGPIBSR
Q.
General
AutoCalibration:EnsuresspecifiedDCandtimingaccuracyismaintainedfor1
yearminimum.
PowerRequirements:Singlephase,100to240V
rms
(±10%)at50/60Hz(±5%);
orsinglephase,100to120V
rms
(±10%)at400Hz(±5%);AutomaticACvoltage
selection
VoltageRange: 90to264V
rms
90to132V
rms
FrequencyRange: 47to63Hz 380to420Hz
PowerConsumption:340Watts(340VA)max‐all4channelmodelsand290
Watts(290VA)‐all2channelmodels,dependingonaccessoriesinstalled
(probes,PCportplugins,etc.);PowerconsumptionduringStandbyState:12
wattsallmodels.
PhysicalDimensions(HWD):260mmx340mmx15
2mm(10.2in.x13.4in.x
6.0in.);heightmeasurementexcludesfootpads
Weight:7.27kg(16lbs.)
Warranty and Service
Threeyearwarranty;calibrationrecommendedyearly.
Optionalserviceprogramsincludeextendedwarranty,upgrades,andcalibration
services.
Environmental Characteristics
TEMPERATURE
Operating:5to40°C
Storage(nonoperating):‐20to+60°C