HUMIDITY
Operating:Maximumrelativehumidity80%fortemperaturesupto31°C
decreasinglinearlyto50%relativehumidityat40°C.
Storage(nonoperating):5to95%RH(noncondensing)astestedperMILPRF
28800F
ALTITUDE
Operating:Upto2,000m
Storage(nonoperating):12,192m(40,000ft)
RANDOMVIBRATION
Operating:5Hzto500Hz,overalllevel:0.31g
rms
,15minutesineachof3
orthogonalaxes
Nonoperating:5Hzto500Hz,overalllevel:2.4g
rms
,15minutesineachof3
orthogonalaxes
SHOCK
FunctionalShock:20gpeak,halfsine,11mspulse,3shocks(positiveand
negative)ineachof3orthogonalaxes,18shockstotal
Certifications
CECompliant,ULandcULListed
CE Declaration of Conformity
TheoscilloscopemeetsrequirementsofEMCDirective89/336/EECfor
ElectromagneticCompatibilityandLowVoltageDirective73/23/EECforProduct
Safety.
EMCDIRECTIVE
EN613261:2006
EMCrequirementsforelectricalequipmentformeasurement,control,and
laboratoryuse.