Viva Series, User - Check & Adjust 633
30.5 Tilting Axis Adjustment (a)
Access In Check & Adjust Wizard, What do you want to do? select Check & adjust the tilting axis
and press Next.
Description This procedure determines the following instrument error:
)
Before determining the tilting axis error, the instrument has to be:
• levelled up using the electronic level
• protected from direct sunlight
• acclimatised to the ambient temperature, approximately 2 minutes per °C difference
compared to the storage place.
• The horizontal collimation error must be determined before.
Check & Adjust Wizard,
Step 1
a Tilting axis error
Key Description
Next To measure the target.
Fn Quit To exit the wizard.