Viva Series, User - Check & Adjust 643
30.8 Configuring Check & Adjust
Access In Check & Adjust Wizard, What do you want to do? select Configure Check & Adjust and
press Next.
Check & Adjust Wizard
Description of options
Next step
Next to change to the Report sheet screen.
Key Description
Next To accept changes and to continue with the subsequent screen
within the wizard.
Back To return to the previous screen.
Fn Quit To exit the wizard.
Option Description
2 weeks, 1 month, 3
months, 6 months
or 12 months
If one or more adjustment values were determined longer ago than
the time specified with this parameter, then a reminder message is
displayed each time the instrument is turned on. This helps to rede-
termine the instrument errors on a regular basis.
Never A reminder message to readjust the instrument is never displayed.
This setting is not recommended.