Page 117
Appendix| Factory Setting
optoNCDT 2300
A 2.2 Set Default Settings
Used hardware:
- PC2300-x/Sub-D
- PC2300-0,5/Y
- RJ45 short-circuit plug
PC2300-0,5/Y
PC2300-x/SUB-D
RJ45
short-circuit
plug
optoNCDT
ERR
Power on
RUN
EtherCAT Ethernet
Laser off
In range
Midrange
Error
PS2020
LASER RADIATION
Do not stare into beam
Class 2 Laser Product
IEC 60825-1: 20xx-xx
P 1mW; P 1.2mW; t=0.5...542 s
0 P
≤ ≤ μ
F=1.5...50kHz;
=670nm
Fig. 61 Default setting with a RJ45 short-circuit plug
Prerequisite: The supply voltage to the sensor is off.
Proceeding:
Connect the RJ45 short-circuit plug on the PC2300-0,5/Y cable, see Fig. 61.
Switch on the supply voltage to the sensor.
Wait until to the end of the boot process in the sensor.
Booting finished
LED Ethernet/EtherCAT yellow
LED State any
Remove the RJ45 short-circuit plug.
i
Resetting the sensor to factory settings with a RJ45 short-circuit plug is possible for sensors that are shipped with a software ver-
sion≥009.xxx.yyy.