Appendix| ASCII Communication with Sensor
optoNCDT 2300
A 6.3.4 Interfaces
A 6.3.4.1 Ethernet
IPCONFIG DHCP|STATIC [<IP address> [<Netmask> [<Gateway>]]]
Set Ethernet interface.
- DHCP: IP address and gateway are automatically requested by DHCP. System looks for a LinkLocal address after appr. 30 seconds
if no DHCP server is available.
- STATIC: Set IP address, net mask, and gateway in format xxx.xxx.xxx.xxx
Values stay the same if no IP address, net mask, and gateway is typed in.
A 6.3.4.2 Setting Measurement Server
MEASTRANSFER NONE|SERVER/TCP [<PORT>]|(CLIENT/TCP|CLIENT/UDP [<IP address> [<Port>]])
The ILD2300 can be operated as a server as well as a client for measurement output via Ethernet.
- None: No measurement transmission via Ethernet.
- SERVER/TCP: Sensor provides a server for the typed in port, under which the measured values can be sent. This is only possible
via TCP/IP.
- CLIENT/TCP: Sensor sends measured values via TCP/IP connection oriented to server.
- CLIENT/UDP: Sensor sends measured values via UDP/IP to server. Therefore the IP address and port on the server are specified.
- IPAddress: IP address of the server, to which measured values are sent when in client-mode, (only valid for CLIENT/TCP or CLIENT/
UDP)
- Port: Port, to which the server gets assigned to in server-mode or to which the measured values are sent in client-mode (min: 1024,
max: 65535)
A 6.3.4.3 Setting RS422
BAUDRATE 9600|115200|230400|460800|691200|921600|1500000|2000000|2500000|
3000000|3500000|4000000
Set the baud rate for the RS422 interface.