Control Menu, Set Sensor Parameter
optoNCDT 2300
7.6.1.2 Signal Processing -
Value Output Trigger
Measurement values are calculat-
ed continuously and independently
of the trigger event. A trigger event
simply triggers the value output via
a digital interface. Therefore, any
values measured immediately be-
fore the trigger event are included
in calculating mean values (aver-
ages) or statistics.
Video signal
Distance calculation
Error identification
Video averaging
Masking
Refractive index correction
Error logging
Found error?
Trigger signal?
no
yes
no
yes
Spike correction
Difference / Thickness
Measurement averaging
Statistics
Hold last value
Relative measurements after
Zeroing / Mastering
Data reduction
GetValue
Data output No data output
Trigger
Trigger processing