Page 97
Value Output
optoNCDT 2300
9.1.2 Error Codes
Error code Description
262073 Scaling error RS422 interface underflow
262074 Scaling error RS422 interface overflow
262075 Too much data for selected baud rate
1
262076 No peak available
262077 Peak is in front of the measuring range
262078 Peak is after the measuring range
262079 Measurement can not be calculated
262080 Measurement can not be evaluated, global error
2
262081 Peak is to wide
262082 Laser beam is off
Fig. 51 Digital error codes RS422
1) This error occurs, if more data should be output, as the may be transmitted with baud rate and measuring rate selected. To fix the
error, there are the following possibilities:
- increase baud rate
- decrease measuring rate
- reduce data amount
if two data words selected, reduce to one data word
reduce output rate
See the following sections:
BAUDRATE, MEASRATE, OUTREDUCE, OUTSTATISTIC_RS422, OUTDIST_RS422, OUTTHICK_RS422
2) This error occurs, if one of the selected peaks is not fully evaluated, because a part of a peak is not recorded. The reasons may be:
- a part of a peak is not recorded by the sensor (before or after the valid measuring range)
- at 49,140 kHz a part of a peak is within the reduced measuring range which is nor used
- when using the ROI a part of a peak is outside the selected range