The LED EtherCAT/Ethernet on sensor is off.
Restart the sensor.
The LED EtherCAT/Ethernet on sensor is yellow.
EtherCAT
Ethernet
A 7.12 Troubleshooting
Initial situation: Sensor erroneously converted to EtherCAT.
Purpose: Enable Ethernet interface.
The TwinCAT-Manager program is installed, the device description file <optoNCDT2300.xml > is copied from the product-CD in the
directory \\TwinCAT\IO\EtherCAT.
Restart the sensor.
Restart the TwinCAT-Manager
Select the menu File > New.
Select the tab I/O Devices, then Scan Devices.
Confirm with OK.
Select a network card, where EtherCAT®–Slaves should be searching for.