EasyManua.ls Logo

MICRO-EPSILON optoNCDT 2300 - Averaging, Error Processing, Spike Correction and Statistics

MICRO-EPSILON optoNCDT 2300
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 59
Control Menu, Set Sensor Parameter
optoNCDT 2300
7.4.4 Averaging, Error Processing, Spike Correction and Statistics
Video averaging No averaging /
Recursive 2 / 4 / 8
Moving 2 / 3 / 4
Median 3
The video averaging is effected before the calculation of
the displacement or thickness. Recommended for very
small peaks respectively to receive more valid data.
Averaging of mea-
surement value
No averaging
Moving N values 2 / 4 / 8 ... 128 Value Indication of averaging mode. The averaging number N
indicates the number of consecutive measurement values
to be averaged in the sensor.
Recursive N values 2 ... 32768 Value
Median N values 3 / 5 / 7 / 9 Value
Error processing Error output, no measurement value Sensor emits error value.
Hold last value 0 ... 1024 Value If no valid measurement value is determined, the last
valid value can be hold for a certain period, that is, output
repeatedly. The last valid value is kept indefinitely at „0“.
Spike correction No
Yes Evaluation length
1 - 10
Value This filter removes individual very high spikes from a rela-
tively constant course of measurement value.
Smaller spikes are preserved.
Max. tolerance range
(mm)
0 - 100
Value
Number of corrected
value 1 - 100
Value
Statistics 2 / 4 / 8 / 16 ... 16384 Beyond a certain number of measurement values the sta-
tistical values minimum, maximum and peak-to-peak are
determined and output.
Grey shaded fields require a selection.
Value
Dark-bordered fields require you to specify a value.