Page 141
Appendix| ASCII Communication with Sensor
optoNCDT 2300
A 6.5 Data Output
A 6.5.1 General
A 6.5.1.1 Selection Digital Output
OUTPUT NONE|RS422|ETHERNET
- NONE: No measurement value
- RS422: Output of measurement values via RS422
- ETHERNET: Output of measurement values via Ethernet
A 6.5.1.2 Output Data Rate
OUTREDUCE <Output reduction> [RS422|ETHERNET|NONE]
Reduces the measurement value output for all available interfaces.
- 1: Output each measurement value
- 2 ... 3000000: Output of each n-th measurement value
A 6.5.1.3 Error Processing
OUTHOLD NONE|0|<Number>
Setting the behavior of the measurement value output in case of error.
- NONE: No holding the last measurement value, output of error value.
- 0: Infinite holding of the last measurement value.
- Number: Holding the last measurement value on the number of measuring cycles; then an error value (maximum of 1024) is output.
A 6.5.1.4 Specified Measured Value Output
GETVALUE NONE|<Number>|ALL
Sends a specified number of measurement value frame. The command works after the commands OUTREDUCE and TRIGGER. It is
no storable parameter. All measurement value frames are always output after Power ON.
- NONE: No measurement value frames are output.
- 1...4294967294: Output of specified number of measurement value frames
- ALL: Continuous output of measurement value frames