Appendix| Control Menu
optoNCDT 2300
A 8 Control Menu
Login, user level
Logged in as Value Read only
Change user level Button Logout or password required
Change password
Old password Value
New password Value
Repeat new password Value
User level upon switching on Professional / User
Measuring
program
Measurement
arrangement
Diffuse reflection /
Direct reflection distance measurement /
Direct reflection thickness measurement
Displacement measurement by diffuse reflec-
tion; Sensor evaluates the reflected stray light.
Displacement or thickness measurement by
direct reflection; Sensor evaluates the light,
which is reflected at the target surface.
Peak to be mea-
sured
first Peak /
highest Peak /
widest Peak
Defines, which signal is used for the evaluation
in the line signal. First Peak: Nearest peak to
sensor. Highest peak: Standard, peak with the
highest intensity.
Widest Peak: Signal with the largest area, use
by small adjacent faults
Laser power full / reduced / off
Reduction of the laser power. Recommended
for strongly reflective targets.
Material Selection from material data base
Material selection necessary for the thickness
measuring program. The factory setting of the
data base can be restored, if the factory setting
is loaded. In the material data base up to 20
materials can be stored.