Appendix| ASCII Communication with Sensor
optoNCDT 2300
A 6.4.4.3 Values used for Statistics
STATISTICDEPTH 2|4|8|16|32|64|128|256|512|1024|2048|4096|8192|16384|ALL
- ALL: Statistics calculated over all values up to the command „RESETSTATISTIC“ is used.
- 2|4|...16384: Range for moving used to calculate the statistics.
A 6.4.4.4 Reset the Statistics
RESETSTATISTIC
Resets the statistical values.
A 6.4.4.5 Setting Masters / Zero
MASTERMV NONE|MASTER <Master value>
- NONE: Completes the mastering.
- MASTER: Set the current measurement value as a master value.
- Master value: Master value in millimeters; min: -2 * measurement value, max: +2 * measuring range.
In case of master value is 0, then the mastering has the same functionality as the zero setting.
The master command awaits the next measurement value a maximum of 2 seconds and masters it. If no measurement value is
received within this time, for example, by external triggering, the command returns with the error “E32 Timeout“. The master value is
processed with six decimal places. Note that the output value is limited to 18 bits during data output via the RS422 interface.
Calculation of a measurement value in mm from digital output:
x [mm]= digital
OUT
- 0.51 Measuring range [mm]
*
*
1.02
65520