Page 67
Control Menu, Set Sensor Parameter
optoNCDT 2300
7.5 Data Output
7.5.1 Digital Interfaces
Selection of digital
interfaces
Web diagram /
Ethernet measurement transmission /
RS422
Decides via the used interface for measurement output.
A parallel measurement output via multiple channels is
not possible.
Data selection Distance / Statistics Min / Statistics Max /
Statistics peak-to-peak / Exposure time /
Intensity of distance value / Status /
Measured value counter / Time stamp /
Trigger counter / Temperature
The data which are provided for the transmission are to
activate with the checkbox.
Settings
Ethernet
IP settings of
the base unit
Address type Static IP address /
DHCP
When using a static IP address, values for IP address,
gateway and subnet mask are to set; and this not applies
when using DHCP.
IP address Value
Gateway Value
Subnet mask Value
Settings of
the Ethernet
measurement
value transmis-
sion
Transmission
type
Client / Server The sensor as server provides the measurement values
at the indicated port or sends them connection-oriented
as client to the indicated client. The measurement value
server is the destination of the measurement values and
can be a PC or a PLC in the network. PC for program-
ming/ demo programming are not measurement value
server.
Protocol PCP/IP / UDP/IP
IP address Value
Port Value Setting the port on the measurement value server;
Settings
RS422
Baud rate 9.6 / 115.2 / 230.4 ... / 4000 kBps Transmission rate with binary data format.
Ethernet/EtherCAT Operation mode after start Ethernet/EtherCAT
Grey shaded fields require a selection.
Value
Dark-bordered fields require you to specify a value.