EasyManua.ls Logo

MICRO-EPSILON optoNCDT 2300 - A 6.4.1.5 Laser Power; A 6.4.2 Video Signal; A 6.4.2.1 Reduction of Region of Interest (ROI); A 6.4.2.2 Video Averaging

MICRO-EPSILON optoNCDT 2300
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 137
Appendix| ASCII Communication with Sensor
optoNCDT 2300
A 6.4.1.5 Laser Power
LASERPOW FULL | REDUCED | OFF
- FULL: Laser power is set to 100 % (1 mW, recommended).
- REDUCED: Laser power is reduced to 10 %; for reflecting materials in direct reflection mode.
- OFF: Laser is switched off.
Switch over of laser power cannot be used for control purposes because switch over is done delayed with low-pass filter. Typically
sensor power is set to 100 %, only on strong reflecting materials (e.g. mirror) a decrease in power is thoughtful.
A 6.4.2 Video Signal
A 6.4.2.1 Reduction of Region of Interest (ROI)
ROI <Start> <End>
Set region of interest. ROI for start and end is between 0 and 511. “Start” value is smaller than “End” value.
A 6.4.2.2 Video Averaging
VSAVERAGE NONE|REC2|REC4|REC8|MOV2|MOV3|MOV4|MED3
- NONE: No video signal averaging
- REC2, REC4, REC8: Recursive average value over 2, 4, or 8 video signals
- MOV2, MOV3, MOV4: Moving average value 2, 4, or 8 video signals
- MED3: Median over 3 video signals