Page 137
Appendix| ASCII Communication with Sensor
optoNCDT 2300
A 6.4.1.5 Laser Power
LASERPOW FULL | REDUCED | OFF
- FULL: Laser power is set to 100 % (1 mW, recommended).
- REDUCED: Laser power is reduced to 10 %; for reflecting materials in direct reflection mode.
- OFF: Laser is switched off.
Switch over of laser power cannot be used for control purposes because switch over is done delayed with low-pass filter. Typically
sensor power is set to 100 %, only on strong reflecting materials (e.g. mirror) a decrease in power is thoughtful.
A 6.4.2 Video Signal
A 6.4.2.1 Reduction of Region of Interest (ROI)
ROI <Start> <End>
Set region of interest. ROI for start and end is between 0 and 511. “Start” value is smaller than “End” value.
A 6.4.2.2 Video Averaging
VSAVERAGE NONE|REC2|REC4|REC8|MOV2|MOV3|MOV4|MED3
- NONE: No video signal averaging
- REC2, REC4, REC8: Recursive average value over 2, 4, or 8 video signals
- MOV2, MOV3, MOV4: Moving average value 2, 4, or 8 video signals
- MED3: Median over 3 video signals