Installation | Housing M
optoNCDT 2300
5.3 Housing M
ILD2300-200 ILD2300-300
ILD2310-10 ILD2310-20 ILD2310-40
Bolt connection Direct fastening
2 Nm
3.5 Nm
8 ... 10
Housing Through
length
Screw Washer Tightening
torque per
screw
Screw depth Screw Tightening
torque per
screw
ISO 4762-A2 ISO 7089-A2 µ = 0.12 Minimum Maximum ISO 4762-A2 µ = 0.12
mm Nm mm mm Nm
M 35 M4 A4.3 2 8 10 M5 3.5
Fig. 10 Mounting conditions housing size M
Recommended tightening torque
max. + 10 % permissible, not deceed min. -20 %!
The tightening torques specified in the table are approximate and may vary depending on the application.
The bearing surfaces surrounding the through-holes (mounting-holes) are slightly raised.
i
Mount the sensor only to the existing through-holes on a flat surface. Clamps of any kind are not permitted.
Housing sizes, see Chap. 3.4.